TYCO TE MEDICAL NELLCOR; OXIMETER Back to Search Results
Model Number PMC10GE360N
Medical Device Problem Code Low Readings (2460)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/02/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that during monitoring a malfunction occurred on the cable, causing the signal to randomly drop to the monitor.The spo2 readings for a patient, who was in the emergency room for a wound on the foot, dropped from 93% to 60%.The cable was replaced with a new one, after which the patient's spo2 readings remained within the normal range.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Additional information: d4, d9, g3, h3, h6 device evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that a malfunction occurred on the cable, causing the signal to randomly drop to the monitor.The spo2 readings for a patient, who was in the emergency room for a wound on the foot, dropped from 93% to 60%.The reported issue was not confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NELLCOR
Common Device Name
OXIMETER
Manufacturer (Section D)
TYCO TE MEDICAL
10025 sw freeman ct
wilsonville OR 97070 9289
Manufacturer (Section G)
TYCO TE MEDICAL
10025 sw freeman ct
wilsonville OR 97070 9289
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key22337141
Report Number2936999-2025-00352
Device Sequence Number13026683
Product Code DQA
UDI-Device Identifier10884521816015
UDI-Public(01)10884521816015(11)231227(21)ST23520103(20)02
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K172482
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPMC10GE360N
Device Catalogue NumberPMC10GE360N
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/02/2025
Supplement Date Received by Manufacturer04/17/2026
Initial Report FDA Received Date06/26/2025
Supplement Report FDA Received Date04/20/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age55 YR
Patient SexFemale
Patient Weight50 KG
-
-