|
Additional information was added to h3, h6 and h11.H11: the device was received for evaluation.Functional testing including leak, clear passage, and clamp function testing were performed.The visual inspection, by the naked eye, along with the functional testing observed a leak from a cut/hole/slice on the front and back side of the bag.The reported condition was verified.The cause was undetermined, however, it is possible a sharp object caused the damage.A batch review was conducted and there were no deviations found related to this condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|