ERIKA DE REYNOSA, S.A. DE C.V. LIBERTY CYCLER SET, SINGLE PATIENT CONN; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number 050-87215A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Peritonitis (2252)
Date of Event 06/01/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Clinical review: there is a possible temporal relationship between pd therapy utilizing the liberty cycler set and the patient event of peritonitis.However, it is unknown if this patient was utilizing fresenius product(s) at the time their reported peritonitis event was diagnosed.There is no patient information and no information to confirm if the patient was a fresenius clinic patient.Furthermore, there is no documentation in the complaint file to show a causal relationship between the peritonitis event and use of the liberty select cycler with cycler set.Additionally, there is no allegation of a device malfunction or deficiency, or cycler set defect reported for the event.All pd patients are at risk of peritonitis due to a compromised immune system and the increase of potential for microbial contamination from dialysis techniques.Most cases of peritonitis are caused by touch contamination by the patient with the pd catheter being a source of entry for organisms into the peritoneum.Based on the limited information and no allegation of a malfunction, deficiency or defect and no confirmation that any fresenius product(s) were utilized, the liberty cycler set can be excluded as the cause of the patient¿s reported peritonitis event.The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.
 
Event or Problem Description
A registered nurse (rn) reported to fresenius that this patient had been hospitalized due to peritonitis.The nurse did not recall any additional information regarding the event or the patient.Attempts to obtain additional information from the acute clinic was unsuccessful.
 
Additional Manufacturer Narrative
Plant investigation: the sample was not returned to the manufacturer and the lot number was not provided.A manufacturing review was performed on the products shipped to the patient for the three (3) month time frame which immediately preceded the event occurrence date.This review included the lot numbers for all fresenius liberty cycler sets shipped to this account within the selected time frame.The entire set of lots have been sold and distributed.There were no non-conformances or abnormalities identified during the manufacturing process which could be associated with the reported event.An investigation of the device history records (dhr) was conducted and confirmed that the results of the in-progress and final quality control (qc) testing met all requirements.The product lots involved met all specifications for release.A review of the dhr did not reveal a probable cause for the customer complaint.As a physical evaluation could not be performed, a definitive conclusion regarding the reported incident could not be reached and a cause could not be confirmed.
 
Event or Problem Description
A registered nurse (rn) reported to fresenius that this patient had been hospitalized due to peritonitis.The nurse did not recall any additional information regarding the event or the patient.Attempts to obtain additional information from the acutes clinic was unsuccessful.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIBERTY CYCLER SET, SINGLE PATIENT CONN
Common Device Name
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
ERIKA DE REYNOSA, S.A. DE C.V.
mike allen #1331
parque industrial reynosa
reynosa 88780
MX  88780
Manufacturer (Section G)
ERIKA DE REYNOSA, S.A. DE C.V.
director, quality systems
900 w zaragosa drive suite d
pharr TX 78577
Manufacturer Contact
jessica trujillo
920 winter st
waltham, MA 02451
6174175172
MDR Report Key22337645
Report Number0008030665-2025-01460
Device Sequence Number15720706
Product Code FKX
UDI-Device Identifier00840861102709
UDI-Public00840861102709
Combination Product (Y/N)N
Initial Reporter StateNM
Initial Reporter CountryUS
PMA/510(K) Number
K173718
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 08/28/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number050-87215A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device AgeMO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/19/2025
Supplement Date Received by Manufacturer08/20/2025
Initial Report FDA Received Date06/26/2025
Supplement Report FDA Received Date08/28/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DELFLEX PD FLUID.; DELFLEX PD FLUID.; LIBERTY SELECT CYCLER.; LIBERTY SELECT CYCLER.
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexUnknown
-
-