• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER ORTHOPAEDIC MFG. LTD. PERSONA TIBIA CEMENTED 5 DEGREE STEMMED RIGHT SIZE D; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER ORTHOPAEDIC MFG. LTD. PERSONA TIBIA CEMENTED 5 DEGREE STEMMED RIGHT SIZE D; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Pain (1994); Aseptic loosening (4883)
Date of Event 06/03/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported a patient underwent a knee arthroplasty on an unknown date.Subsequently, the patient was revised on due to loosening.Attempts have been made and all available information has been provided.
 
Additional Manufacturer Narrative
(b)(4).G2: foreign: australia.D10: unk; femoral component; unk.Unk; unk articular surface; unk.Customer has indicated that the product will not be returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: a2; a3; b4; b5; d1; d2; d4; d6; d10; g1; g3; g4; g6; h1; h2; h4; h6.D10: cat # 42502805802, femur trabecular metal cruciate retaining (cr) standard porous, 64709072 unknown catalog, articular surface, unknown lot.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported a patient underwent a knee arthroplasty.Subsequently, the patient experienced pain and was revised due to loosening.Attempts have been made and all available information has been provided.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4; b5; d2; g1; g3; g6; h1; h2; h3; h6.The event cannot be confirmed.No product was returned, or pictures provided; visual and dimensional evaluations could not be performed.The device history record was reviewed and no discrepancies relevant to the reported event were found.Review of complaint history found no additional related issues for this item and the reported part and lot combination.An x-ray was provided and assessed.The x-ray was not sent for review as a single undated image would not provide an accurate assessment of loosening which was reported to have occurred within 4 months of implantation.Sending the image would not enhance the investigation.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERSONA TIBIA CEMENTED 5 DEGREE STEMMED RIGHT SIZE D
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
Manufacturer (Section G)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare V14 E 370
EI   V14 E370
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key22338032
Report Number3007963827-2025-00255
Device Sequence Number21073516
Product Code MBH
UDI-Device Identifier00889024246683
UDI-Public(01)00889024246683(17)321024(10)65516786
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K172524
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/21/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42532006702
Device Lot Number65516786
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/03/2025
Supplement Date Received by Manufacturer06/26/2025
08/26/2025
Initial Report FDA Received Date06/26/2025
Supplement Report FDA Received Date07/22/2025
09/22/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/09/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11.; SEE H11.
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexFemale
-
-