| Catalog Number |
367863 |
| Medical Device Problem Code |
Material Protrusion/Extrusion (2979)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/30/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported prior to using bd vacutainer® k2 edta (k2e) 10.8mg blood collection tubes, there were 50 hemogard closures with a sharp burr on them.No adverse impact was reported regarding the patient or user.
|
| |
|
Additional Manufacturer Narrative
|
|
E1: initial reporter facility name: (b)(6).There were multiple medical device types: d2b: medical device type: jka there were multiple 510k numbers g4: pma / 510(k)#: k213670, k231373 a device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported prior to using bd vacutainer® k2 edta (k2e) 10.8mg blood collection tubes, there were 50 hemogard closures with a sharp burr on them.No adverse impact was reported regarding the patient or user.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary bd received one photo for investigation, which shows a piece of plastic protruding from the gate on the hemogard closure.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.This complaint has been confirmed for the indicated failure mode: molding defect.No definitive root cause could be established for the reported defect.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for the identification of emerging trends.
|
| |
|