| Model Number |
G18343 |
| Medical Device Problem Codes |
Use of Device Problem (1670); Premature Separation (4045)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/12/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During a colonoscopy, the physician used a cook instinct endoscopic hemoclip.It was reported that the clip was released prematurely by removing it from the sterile packaging and attempting to position it in the working channel of the endoscope, without putting pressure on the activation mechanism [premature deployment in the scope].An unintended section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.The instructions for use states: "uncoil device.Verify smooth handle operation and clip action.Open clip by gently moving handle spool distally (away from handle thumb ring).Once clip is fully open, do not continue advancing handle spool as clip may prematurely detach from catheter.Close clip by moving handle spool proximally until clip is fully closed.Caution: do not continue to pull handle spool beyond tactile resistance as this may prematurely deploy clip." the instructions for use includes the following precaution: "endoscope must remain as straight as possible when inserting or withdrawing device." the instructions for use also states: "with clip closed and without holding handle spool, advance device in small increments into accessory channel of gastroscope or colonoscope." failure to follow the instructions above can result in damage to the device which may lead to premature clip deployment.Prior to distribution, all instinct endoscopic hemoclips are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: based on the quality engineering risk assessment no corrective action is warranted at this time.A review of the complaint history was conducted.Based on this review, the likelihood of this type of report is rare.Quality assurance will continue to monitor for complaint trends.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: the additional information indicated the user held the handle spool during advancement through the accessory channel.This is the most likely cause of this report.The instructions for use states: "with clip closed and without holding handle spool, advance device in small increments into accessory channel of gastroscope or colonoscope.Caution: holding handle spool during advancement may prematurely deploy clip." failure to follow these instructions can result in damage to the device which may lead to premature clip deployment.However, even if the clip is not deployed, damage can occur to internal device components such as the driver legs.Once this damage occurs, the clip is in a partially deployed state and may not be able to be opened/reopened.Prior to distribution, all instinct endoscopic hemoclips are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: based on the quality engineering risk assessment no corrective action is warranted at this time.A review of the complaint history was conducted.Based on this review, the likelihood of this type of report is rare.Quality assurance will continue to monitor for complaint trends.Additional comments: based on the information provided that the user held the handle spool during advancement through the accessory channel, a cook representative has been directed to contact the medical facility involved in an effort to promote further education and understanding related to appropriate usage of this product.
|
| |
|
Event or Problem Description
|
|
During a colonoscopy, the physician used a cook instinct endoscopic hemoclip.It was reported that the clip was released prematurely by removing it from the sterile packaging and attempting to position it in the working channel of the endoscope, without putting pressure on the activation mechanism.Additional information received on 26aug2025 indicated that the clip tromboned [clip housing detaches from the cath attach and remains attached to the drive wire].An unintended section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|
| |
|