| Catalog Number |
550000360 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Pain (1994); Joint Laxity (4526)
|
| Date of Event |
06/09/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that patient had shoulder done originally on (b)(6) 2024.Patient was complaining of some pain associated with impingement of the conjoint tendon.The surgeon felt the patient would benefit from an exchange of the glenosphere, humeral shell, and poly.A 32+4 glenosphere was removed and replaced with a 36+4 eccentric.The eccentricity was dialed posterior.A new shell and poly were trialed and the shoulder was found to be stable with improved range of motion.The real shell and poly were opened and implanted and stability was confirmed once reduced.No delays or adverse events took place.Activity level - yes.Doi: (b)(6) 2024.Dor: unknown.Affected side: left shoulder.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not perform as part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.H11 additional narrative: added: d10.
|
| |
|
Search Alerts/Recalls
|