| Catalog Number |
EIC8872-01 |
| Medical Device Problem Code |
Melted (1385)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/02/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).This complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Event or Problem Description
|
|
It was reported that during an ent procedure, the metal electrode wire of the procise xp fused.The procedure was successfully completed with a delay of greater than 30 minutes using a smith and nephew back up device.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: h3, h6: the reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.However, an analysis of the customer provided image found the device is being held at the tip showing there are no electrodes left.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.Factors that could have contributed to the failure include activating the device above recommended settings for prolonged periods of time.Please refer to the instructions for use for precautions and warnings related to the use of the device.Based on this investigation, the need for corrective action is not indicated.If any additional information related to this event is received in the future, this investigation will be reopened for further assessment.
|
| |
|
Additional Manufacturer Narrative
|
|
H2: corrected information in h6 (type of investigation & investigation findings).H3, h6: the reported device was received for evaluation.Visual inspection observed the returned instrument shows no manufacturing abnormalities.Product was out of the original packaging.No packaging returned.The tip is worn from use.The electrodes are thin from use.Only a small portion of the center and bottom electrode remain.The top electrode is thin but has most of its body.The opened device was tested and plugged into the controller and registered settings (7,3).The wand produced plasma with its remaining electrode and had no issues activating coag.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A definitive root cause for the reported issue could not be determined.Factors that could have contributed to the failure include procedural variance, or deviations from established protocols, increasing risks and leading to unintended consequences (1) hitting the device tip against a hard surface.2) using the device as a lever to enlarge surgical site.3) mechanical displacement of tissue through applied force.4) activating the device above recommended settings for prolonged periods of time).Based on this investigation, the need for corrective action is not indicated.
|
| |
|