| Model Number |
630-2131111-01 |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation for evaluation.During disassembly of the device to determine root cause, the technician observed damage consistent with mis-handling.The lcd was replaced.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during functional testing, the device's screen turned completely white and was illegible.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|