|
During the atrial fibrillation procedure, the sheath was inserted into the heart.After the viewflex catheter was inserted and then removed, blood leaked from the hemostatic valve when no catheter was inserted.The sheath was replaced, and the procedure was completed with no adverse consequences to the patient.No air movement into the patient was confirmed.The only products inserted directly into the hemostasis valve were the included dilator and viewflex catheter.No additional venipuncture was performed due to sheath replacement.There was no resistance when inserting the dilator.The sheath and dilator were integrated and inserted as per the procedure.
|
|
Additional information: d9, g3, g6, h2, h3, h6, h11.One 10f fast-cath introducer sheath was received for evaluation.The sheath passed pressure and aspiration leak testing with no anomalies observed.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported leak remains unknown.
|