| Model Number |
60005 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Visual Disturbances (2140)
|
| Date of Event |
06/02/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device history record for the manufacturing lot was reviewed and there were no discrepancies or unusual findings.Manufacturer reference #:(b)(4).
|
| |
|
Event or Problem Description
|
|
A site reported to rxsight that a light adjustable lens (lal, 70057600118, +20.0d) was explanted due to visual disturbances.A new lal was implanted as a replacement.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide additional information to the following sections: d9, h3, h6, and h11.The lal was cut into two pieces during explantation.No device problem was found during visual inspection.Manufacturer reference #: (b)(4).
|
| |
|
Search Alerts/Recalls
|