Model Number RFG-IONIC
Medical Device Problem Code Temperature Problem (3022)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/02/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an rfa procedure, multiple probes were not reaching temperature.Troubleshooting was done but the issue persisted.The procedure was abandoned.There were no adverse patient consequences.
 
Additional Manufacturer Narrative
The reported event was confirmed.Review of the log files confirmed a temperature rise time warning on channel on multiple channels; however the device functioned normally throughout product testing.No non-conformances associated with the reported event were identified.Based on the information received and the investigation performed, the cause for the reported event was unable to be determined.The returned product functioned properly during the evaluation.No hardware abnormalities that would have resulted in the reported event were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IONICRF¿ GENERATOR
Common Device Name
RFA GENERATOR
Manufacturer (Section D)
ABBOTT MEDICAL (AF-ST. PAUL)
one st. jude medical drive
st. paul MN 55117
Manufacturer (Section G)
ABBOTT MEDICAL (AF-ST. PAUL)
one st. jude medical drive
st. paul MN 55117
Manufacturer Contact
heidi syndergaard
6901 preston road
plano, TX 75024
9723098000
MDR Report Key22342816
Report Number2184149-2025-00122
Device Sequence Number18285975
Product Code GXD
UDI-Device Identifier05415067029317
UDI-Public(01)05415067029317(10)10701393
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberRFG-IONIC
Device Catalogue NumberRFG-IONIC
Device Lot Number10701393
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/03/2025
Supplement Date Received by Manufacturer07/07/2025
Initial Report FDA Received Date06/26/2025
Supplement Report FDA Received Date07/10/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/04/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age33 YR
Patient SexFemale
Patient Weight84 KG
-
-