• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DORNIER MEDTECH GMBH COMPACT SIGMA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DORNIER MEDTECH GMBH COMPACT SIGMA Back to Search Results
Model Number K1025660
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Hematoma (1884)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This complaint was documented subsequent to the initial reporting.A service report was completed by dmta field service engineer following the evaluation of the device identified by this complaint.This technical system safety check was conducted at 22 june 2025 and concluded that the device was in compliance with dornier specification.Patient hematoma is listed as potential adverse effect and complication in the sigma operating manual.Details concerning the patient outside of the confirmation of the hematoma were not provided to dmta for review.The result of this investigation revealed no defects or inconsistencies with the identified device.The reviews conducted for this investigation concluded the device was manufactured and functioning within dornier specifications.
 
Event or Problem Description
Patient hematoma was reported and initially reported to fda under 1037955-2025-00025 on 06/26/2025.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMPACT SIGMA
Common Device Name
SIGMA
Manufacturer (Section D)
DORNIER MEDTECH GMBH
argelsrieder feld 7
wessling, bavaria 82234
GM  82234
Manufacturer (Section G)
DORNIER MEDTECH GMBH
argelsrieder feld 7
wessling, bavaria 82234
GM   82234
Manufacturer Contact
stefan lutzenberger
argelsrieder feld 7
wessling, bavaria 82234
GM   82234
MDR Report Key22343279
Report Number9611388-2025-00017
Device Sequence Number20033567
Product Code LNS
UDI-Device Identifier04049957000011
UDI-Public(01)04049957000011
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K002929
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 06/27/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberK1025660
Device Catalogue NumberK1025660
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/24/2025
Initial Report FDA Received Date06/27/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/09/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-