BOSTON SCIENTIFIC CORPORATION INOGEN? EL ICD DR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number D143
Medical Device Problem Code Impedance Problem (2950)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/04/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If pertinent information is provided in the future, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that this implantable cardioverter defibrillator (icd) system exhibited out of range right atrial and right ventricular pacing impedance measurements.The health care professional reported isometrics were completed and no oversensing was observed.Additional information has been requested but was not provided at this time.This icd remains in service.No adverse patient effects were reported.
 
Additional Manufacturer Narrative
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that this device exhibited intermittent changes in impedance measurements with no conclusive evidence of a product performance issue or inadequate lead-to-device connection; please refer to the b5, describe event or problem field for more information regarding the specific circumstances of this event.Engineering analysis and testing of returned products has determined that repeated, small movements of the lead terminal ring along with variances in the connection contact force and surface roughness can impact the connection between the spring contact and the lead ring, resulting in intermittent changes in impedance measurement.If pertinent information is provided in the future, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that this implantable cardioverter defibrillator (icd) system exhibited out of range right atrial and right ventricular pacing impedance measurements.The health care professional reported isometrics were completed and no oversensing was observed.Additional information was requested but was not provided.Due diligence has been completed.This icd remains in service.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INOGEN? EL ICD DR
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key22346858
Report Number2124215-2025-42870
Device Sequence Number8573344
Product Code LWS
UDI-Device Identifier00802526534263
UDI-Public(01)00802526534263(17)171214
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P960040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/20/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date12/14/2017
Device Model NumberD143
Device Catalogue NumberD143
Device Lot Number203104
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/23/2025
Supplement Date Received by Manufacturer08/20/2025
Initial Report FDA Received Date06/27/2025
Supplement Report FDA Received Date08/20/2025
Date Device Manufactured01/12/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age59 YR
Patient SexFemale
-
-