| Model Number |
N/A |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/06/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The date of event was not provided by the complainant/reporter, the date reflected in this report is the date bd became aware of the event.The manufacturer has received the sample and will be evaluated.A supplemental will be submitted with evaluation results.
|
| |
|
Event or Problem Description
|
|
It was reported, picc placed on 5/30 - lasted three days.Could not flush.No blood return.Put in cathflo, didn't work.Did an x-ray of arm and kinked in arm 5 cm from 0 mark.Patient needed x-ray and new picc placement.
|
| |
|
Event or Problem Description
|
|
Additional information provided 08-jun-2025: admit date: (b)(6) 2025.Picc length: 39cm.Patient's arm circumference: 30.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The complaint of a kinked catheter was confirmed.The product returned for evaluation was one 4 fr s/l powerpicc.The investigation findings are consistent with damage accumulated through flexural fatigue.Flexural fatigue may occur due to cyclic kinking of the catheter tube in which physiological, placement, usage, and mechanical factors may gradually form a crack(s) in the catheter.The returned product sample was evaluated, and a kink was observed between the 3 cm and 4 cm depth markings.The catheter kink contained physical features associated with material fatigue, and the characteristics observed which supported this type of failure included: ¿ damage which was circumferentially aligned.¿ 'c' shaped impressions leading into the kink site which are consistent with material buckling due to movement which caused the fracture edges to be pressed together.¿ overall elliptical shape to the kink cross-section (a result of repeated kinking of the tubing).An examination of the catheter structure revealed no potential damage/defect related to manufacture of the product.The damage location suggested that catheter securement, access and maintenance techniques may have contributed.This complaint will be recorded for future trending and monitoring purposes.
|
| |
|