• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOTRONIK SE & CO. KG ENTOVIS DR-T PROMRI; PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK SE & CO. KG ENTOVIS DR-T PROMRI; PACEMAKER Back to Search Results
Model Number 371992
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Abscess (1690); Endocarditis (1834); Bacteremia (4846)
Date of Event 06/20/2024
Type of Reportable Event Serious Injury
Event or Problem Description
System was explanted due to bacteremia, endocarditis, and a spinal abscess.Should additional information become available, this file will be updated.
 
Additional Manufacturer Narrative
An infection was observed following the implantation of this biotronik device.The sterilization process was investigated.The validated process assures that all sterilization parameters, such as gas concentration, temperature, humidity, etc., are within its specified ranges for each distributed device.Additionally, an analysis of validated microbiological indicators is performed after every sterilization procedure as evidence of successful completion of the sterilization process.Review of the biotronik complaint database did not reveal any changes regarding the trend for this type of incident.In summary, the infection was not device related.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTOVIS DR-T PROMRI
Common Device Name
PACEMAKER
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
DE  12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key22350262
Report Number1028232-2025-03096
Device Sequence Number8573528
Product Code NVZ
UDI-Device Identifier04035479115173
UDI-Public(01)04035479115173(17)160930
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
P950037
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/27/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Expiration Date09/30/2016
Device Model Number371992
Device Catalogue Number371992
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/02/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/09/2025
Initial Report FDA Received Date06/27/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/27/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Age84 YR
Patient SexFemale
-
-