| Catalog Number |
ES8918 |
| Medical Device Problem Code |
Failure to Deflate (4060)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/12/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The lot number was reviewed for complaint trend, nonconforming report and capa.Devices met specification prior to release.No trends were noted for complaints and there were no nonconforming reports or capas that were confirmed to be associated.
|
| |
|
Event or Problem Description
|
|
According to the available information during the implant procedure, one of the cylinders would not drain during the draining process.The cylinders were removed and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Titan pump and cylinders 1 and 2 were received for evaluation.No functional abnormalities were noted with the pump.No functional abnormalities were noted with cylinder 1 or cylinder 2.The information received indicated the device was not able to deflate, but because no functional abnormalities were noted with the returned components, the complaint could not be confirmed as reported.A review of the device history record confirmed the devices from this lot met all specifications prior to release.A review of the complaint history database, nonconformances and capas revealed no trends for this lot.
|
| |
|
Event or Problem Description
|
|
According to the available information during the implant procedure, one of the cylinders would not drain during the draining process.The cylinders were removed and replaced.
|
| |
|
Search Alerts/Recalls
|