| Model Number |
KD-VC631Q-07201S |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/04/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported during a therapeutic endoscopic retrograde cholangiopancreatography (ercp) procedure, the cutting wire broke.The procedure was completed with an olympus back up device.There were no report patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned to olympus for evaluation/inspection.Based on the results of the investigation, the cutting wire broke could not be identified.A definitive root cause could not be established.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
No additional information received from the customer.
|
| |
|
Search Alerts/Recalls
|