MENTOR TEXAS MENTOR MEMORYGEL XTRA BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Catalog Number |
SMHX375 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/06/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.Since no lot number was provided, no manufacturing record evaluation review could be performed.D4: udi: as the catalog/model number was not provided, the (01) gtin is not available.D4: udi: as the lot, serial number, and the corresponding expiration date for the device involved in the event was not provided, the full udi is currently not available.Reason for device explant and/or reoperation: na.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that an unknown gel implant being used for a breast surgery ruptured intraoperatively.There were no patient consequences or additional information reported.
|
| |
|
Additional Manufacturer Narrative
|
|
On july 14, 2025, the mentor failure analysis lab received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.As per device, and paperwork received, following information has been updated on this form: event day from 1 to 6 under field b3, field d1 for brand name has been updated to "mentor memorygel xtra breast implant," field d4 for catalog number has been updated to "smhx375", field d4 for lot number has been updated to "2044779", field d4 for serial number has been updated to (b)(6), field d4 for unique identifier(udi) has been updated to "(b)(4).A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On august 5, 2025, device evaluation was completed as follows: mentor conducted a visual inspection, microscopic examination, and thickness measurement of the returned device.Visual analysis of the returned device revealed that the breast implant was found to have a tear on the anterior view, measuring approximately 7.0 cm.Microscopic examination was performed, and the cause of the tear could not be identified.Therefore a thickness measurement was conducted on the shell at the tear, and the thickness was within manufacturing specifications.Mentor also performed a manufacturing record evaluation related to the reported complaint for the finished device lot number.No manufacturing non-conformances were identified as part of this evaluation.Although no conclusion could be reach on the cause of the reported event, the instructions for use contain the following caution: according to the product insert data sheet, it is possible to damage the implant during insertion.Ensure that excessive force is not applied to a very small area of the shell during insertion of the device through the incision.Instead, apply force over as large an area of the implant as possible.Avoid pushing the device into place with one or two fingers in a localized area, as this may create an area of weakness on the shell.In addition, an incision should be of appropriate length to accommodate the style, size, and profile of the implant.This will reduce the potential for creating excessive stress to the implant during insertion, and will avoid the risk of damage to the implant.As part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria: no corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.Manufacturer¿s reference number: (b)(4).
|
| |
|