|
(b)(4) date sent: 06/29/2025 d4: batch #397d29 and u5dn7p.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the tr45g reload was received loaded in an ats45 device with no apparent damage.The reload was received partially fired 1/10 and with the reload lockout spring damaged.In addition, 3 unformed staples were received inside a petri dish.The returned device was tested for functionality with test reload and it fired, cut, and formed all the staples as intended.The staple line and cut line were complete, and the staples met the staple form release criteria.The event reported was confirmed and it is related to improper use of the device.The damage to the reload lockout spring is consistent with damage observed when the firing cycle is started, interrupted, released, and restarted.When firing the device make sure that the firing stroke is completed.Do not partially fire the device.Fire the device by squeezing the firing trigger completely until it rests on the closing trigger.Once the firing cycle has been initiated, it must be completed.If the re-initiation of firing is resumed, the device will lockout.Firing through the lockout mechanism will break the device.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished tr45g, with lot/batch number u40l2m, and no non-conformances were identified.A manufacturing record evaluation was performed for the finished device and reloads batch numbers (397d29 and u5dn7p), and no non-conformances related to the reported complaint condition were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
It was reported that preoperatively to an unknown procedure, it was observed that the stapler fire locked out prior to use.The trigger was squeezed prior to fire when inserted in trocar, accidental.There was no patient consequence nor surgical delay reported.No additional information provided.
|