|
(b)(4).Date sent: 06/29/2025.D4: batch #892c02.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that one pce45a device was returned with no apparent damage and with no reload present.The device was tested for functionality in the straight position with a test reload and achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the staples meet the staple release criteria.During the functional test, it was noticed that the firing speed was abnormally slow.In order to verify the condition of the internal components, the device was disassembled.Upon disassembling, the pcb board was noted to be malfunctioning.No conclusion could be reached as to what may have caused the pcb board to be damaged.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished pce45a, with lot/batch number c9e901, and no non-conformances were identified.A manufacturing record evaluation was performed for the finished device batch number 892c02, and no non-conformances related to the reported complaint condition were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|