DEPUY MITEK LLC US GRYPHON P BR ANCHOR W/OC; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
|
Back to Search Results |
|
| Catalog Number |
210814 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11 additional narrative: d4 h4: the device expiration, complete udi and date of manufacture are unknown at this time.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.
|
| |
|
Event or Problem Description
|
|
It was reported that during an unknown surgery, it was observed that the gryphon p br anchor w/oc device anchor was broken.It was reported that all the broken parts were removed.Another device was used to complete the surgery.There were no adverse consequences to the patient.No additional information could be provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11 additional narrative: d9, h3, h6: the actual device has been returned and is currently pending evaluation.Once the device has been evaluated, a supplemental medwatch report will be sent accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11 additional narrative: device history review: there was a non-conformance not related.D4: the device expiration date was unknown in the initial report.The date of expiration has been updated accordingly.H4: the device manufacture date was reported as unknown in the initial report.Please note that the date of manufacture has been updated accordingly.The product was returned to j&j medtech orthopaedics for evaluation.Visual inspection of the returned device found the anchor broken.Broken fragment, suture and needles were returned for evaluation.No other anomalies were noted.The overall complaint was confirmed as the observed condition of the gryphon p br anchor w/oc would have contributed to the complained device issue.Based on the investigation findings, the root cause is traced to procedural variables, such handling of the device or product interaction during procedure; where an axial misalignment may had occurred, that subsequently led to an anchor breakage.As per instructions for use 109363, axial misalignment or levering with the anchor upon insertion, may result in anchor or inserter fracture.Properly handling and attention to the approved use of the device diminishes the risk of failure.It has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11 additional narrative: d4: udi updated from (b)(4) incomplete to (b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|