| Model Number |
VOLAR DISTAL RAD PLT TI STD LT 5H |
| Medical Device Problem Code |
Defective Device (2588)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
06/04/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Event or Problem Description
|
|
On 6/4/2025, a sales representative reported via phone that an ar-8724v-24 2.4mm distal locking screw failed to lock to the ar-8916vsl-05 volar distal radius plate, instead threading through it.The faulty screw was removed, and a new screw was successfully implanted in the central hole, achieving proper locking with the plate.Subsequently, the surgeon implanted the originally faulty screw in a different hole, where it threaded in without any issues.This was discovered during a wrist fx procedure.No additional information is known regarding this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error of the device due to applying excessive mechanical forces during screw insertion and/or over-torquing the screw during insertion.
|
| |
|