ZIMMER ORTHOPAEDIC MFG. LTD. PERSONA ARTICULAR SURFACE MEDIAL CONGRUENT (MC) LEFT 10 MM HEIGHT; PROSTHESIS, KNEE Back to Search Results
Catalog Number 42512100710
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Joint Laxity (4526)
Date of Event 06/11/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).D10: unknown - unknown femoral component - unknown.Unknown - unknown tibial component - unknown.G2: foreign country: australia customer has indicated that the product will not be returned to zimmer biomet for investigation as it was discarded.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that the patient underwent an initial left knee surgery.Approximately 13 months post-op, the patient reported that their knee felt loose and the surgeon suspected that the poly was not thick enough.Subsequently, the patient was revised due to instability.The poly was revised and increased to a larger thickness.There were no signs of poly wear.Attempts have been made and all available information has been provided.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4; b5; d2; g1; g3; g6; h1; h2; h3; h4; h6 the event cannot be confirmed.No product was returned, or pictures provided; visual and dimensional evaluations could not be performed.The device history record was reviewed and no discrepancies relevant to the reported event were found.Review of complaint history found no additional related issues for this item and the reported part and lot combination.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERSONA ARTICULAR SURFACE MEDIAL CONGRUENT (MC) LEFT 10 MM HEIGHT
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
Manufacturer (Section G)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI  
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key22357035
Report Number3007963827-2025-00261
Device Sequence Number8759985
Product Code MBH
UDI-Device Identifier00889024469211
UDI-Public(01)00889024469211(17)280621(10)65873574
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K150090
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/30/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue Number42512100710
Device Lot Number65873574
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/11/2025
Supplement Date Received by Manufacturer07/28/2025
Initial Report FDA Received Date06/30/2025
Supplement Report FDA Received Date07/30/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/26/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11.
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexUnknown
-
-