| Catalog Number |
SRX-340 |
| Medical Device Problem Codes |
Device Appears to Trigger Rejection (1524); Material Rupture (1546)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported a right side capsular contracture, baker grade iii.Healthcare professional also reported ¿damage caused by explant¿ as ¿hole on the bottom¿.The device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been/will be requested.No additional information is available at this time.Reason for reoperation: capsular contracture, baker grade iii; "hole on the bottom".
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a right side capsular contracture, baker grade iii.Healthcare professional also reported ¿damage caused by explant¿ as ¿hole on the bottom¿.The device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the product analysis performed, the assessments of the complaints are: rupture: observed broken device assessed as surgical damage and missing piece of shell assessed as inconclusive.Capsular contracture: unable to observe as it is not related to the manufacturing process.As per the investigation procedures, creases were completed and none of the observations are found to be potentially related to the manufacturing process, no further actions are required.Additional, changed, and/or corrected data: d9, h3, h6.
|
| |
|