This report is being submitted as part of remediation activities associated with capa 643143, addressing mdr reporting guidance from the fda¿s 1995 preamble, which ensures complaints related to corrections and recalls previously reported to the fda under 21 cfr 806 are now reported as mdrs; this is not a new malfunction/event.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Medtronic received information that during the implant of this transcatheter bioprosthetic valve, after one recapture due to the valve position being too high, the handle felt loose.The valve was recaptured multiple times and on the final recapture the handle of the delivery catheter system (dcs) came apart when the capsule was 80% closed.The physician was able to force the handle closed bypushing hard on the clear slider.The dcs was recovered and a new system was used.No adverse patient effects were reported.
|