|
Lot number of the device not provided by the complainant; therefore, the udi, expiration and manufacturing dates are not known.Device history record (dhr) review was unable to be conducted for the disposable device as the identification numbers were not provided by the complainant.The device involved in this event was not returned for evaluation purposes therefore visual and functional analysis of the product could not be performed.We are unable to confirm a relationship between the device and the issue reported.The information obtained during complaint investigation will be included in our global complaint trending and product surveillance will continue to monitor complaints of this type for adverse trends.If the product is received or additional information is obtained, the investigation will be reopened accordingly per standard operating procedure.
|
|
It was reported that during an atec procedure on april 10th, 2025, an artifact was observed in the patient´s breast after an mri biopsy in addition to the marker (trimark marker) that was deployed at the time.The patient received another mammogram on april 10th, 2025, due to the artifact and the radiologist recommended that the patient have surgery to have the material removed, which was scheduled for (b)(6) 2025.No issues were reported during the device set up.No further information is available at this time.
|
|
An investigation was conducted: it was reported that during an atec procedure on (b)(6) 2025, an artifact resembling metal was observed in the patient´s breast after an mri biopsy in addition to the marker (trimark marker) that was deployed at the time.The biopsy was benign, and the radiologist recommended the patient have surgery to have the material removed.No issues were reported during set up and the patient received another mammogram on (b)(6) 2025, due to artifact.Surgery is scheduled for (b)(6) 2025.The device was not available for return; visual and functional analysis/testing could not be conducted for the described event.The device was not returned for this complaint, and only limited patient-related information was provided.Based on the available information, the customer reported material found in the patient¿s breast after biopsy, which was considered relevant to the failure mode.Investigation of this issue was conducted through customer-provided information, historical complaint review, analysis of similar events, and evaluation of product design.Cause/root cause: the root cause of the reported issue is associated with inadequate manufacturing process controls at the supplier, which resulted in introducer sheath components being produced with internal burrs and residual debris.These imperfections were not effectively detected during the previous inspection process, allowing foreign particles to remain in the components and subsequently be present in the final assembled product.A detailed review of historical complaints and nonconformance events confirmed that the debris found during internal inspections matched the particles described in multiple complaints, establishing a direct link between the supplier¿s process deficiencies and the reported failure mode.To address this issue, a scar was issued to the supplier, requiring the implementation of enhanced process controls, burr removal procedures, and improved cleaning steps at the supplier¿s facility.Initially, the complaint from the customer was filed against the atec biopsy device and marker.However, following investigation, it was determined that particles resembling those described in the customer complaint were found inside the introducer's sheath.As the customer confirmed that the atec introducer localization system (ils) was used during the procedure, it has been determined that the foreign material originated from the atec ils and mdr 1222780-2025-00358 was filed against the accessory.Conclusion: the reported issues have been confirmed, and the most probable root cause has been identified.A health risk assessment will establish the need for updates to the risk management.Additionally, an nce was escalated to a scar and through this scar, containment and corrective actions were implemented by the supplier, and the root cause was identified for further investigation and resolution.A capa nor a new hra, scar or nce are deemed required by this event.Future events will be monitored and trended.Complaint confirmed: yes device history record (dhr) review: the dhr review was not conducted since the lot/serial number was not provided by the complainant.
|