OLYMPUS WINTER & IBE GMBH SHAFT "HIQ+", BIPOLAR, 5 X 330 MM, ISOLATED; WA60800C Back to Search Results
Model Number WA60800C
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/05/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
 
Event or Problem Description
It was reported the surgical shaft tip of the flat bipolar was observed to be burnt.The issue was observed during an unspecified therapeutic procedure.There were no reports of patient harm.
 
Additional Manufacturer Narrative
The device was returned to olympus for inspection and the reported failure was not confirmed.A root cause could not be identified.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
It was reported the surgical shaft tip of the flat bipolar was observed to be burnt.The issue was observed during an unspecified therapeutic procedure.There were no reports of patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHAFT "HIQ+", BIPOLAR, 5 X 330 MM, ISOLATED
Common Device Name
WA60800C
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM   22045
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key22365638
Report Number9610773-2025-04237
Device Sequence Number20935339
Product Code GEI
UDI-Device Identifier04042761062398
UDI-Public04042761062398
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K984417
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 09/29/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberWA60800C
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 06/05/2025
Supplement Date Received by Manufacturer09/12/2025
Initial Report FDA Received Date06/30/2025
Supplement Report FDA Received Date09/29/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-