Model Number PED2-375-12
Medical Device Problem Code Retraction Problem (1536)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/25/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a re-treatment of an unruptured, saccular vertebral aneurysm located in the right vertebra with a maximum diameter of 7 mm and a neck diameter of 4 mm, the pipeline flex shield 3.75mm x 12mm was used.After deployment, the pipeline delivery catheter would not re-enter the phenom 27 catheter, and the ptfe sleeves would not enter the phenom 27, necessitating removal of the system as a unit, which was not the preferred method.The vessel tortuosity was described as moderate.Dual antiplatelet treatment was administered, and the platelet reactivity unit level was measured at 220.The angiographic result post-procedure was reported as good.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Product analysis: ¿ as found condition: the pushwire was returned stuck at the distal tip of the phenom 27 catheter, inside a bio-hazard bag, and a shipping box.The pipeline flex shield braid was not returned as it was implanted in the patient.¿ damage location details: the distal and proximal dps restraints were intact.The dps sleeves showed no signs of damage and were found to be intact.The distal hypotube and ptfe shrink tubing were also intact, exhibiting no signs of elongation.However, the distal pushwire appeared to be kinked approximately 1.8 cm from the distal tip coil.No defects were found in the tip coil, distal marker, re-sheathing marker, resheathing pad, or proximal bumper.An examination of the catheter tip and marker revealed no damage.The catheter body was flattened between 2.0 cm and 7.6 cm from the distal tip.No other anomalies were noted.¿ testing/analysis: the pushwire was difficult to push out from the catheter lumen.The total and usable lengths were measured and found to be within specifications.The catheter was flushed with water and was found to be patent.An in-house 0.026¿ mandrel was then tested by running it through the catheter hub, successfully passing through without issues.However, it became stuck at 7.6 cm from the distal tip.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PIPELINE FLEX W/SHIELD TECHNOLOGY
Common Device Name
INTRACRANIAL ANEURYSM FLOW DIVERTER
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer (Section G)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer Contact
justin ellis
9775 toledo way
irvine, CA 92618
7635265677
MDR Report Key22365832
Report Number2029214-2025-01488
Device Sequence Number18250088
Product Code OUT
UDI-Device Identifier00763000284237
UDI-Public(01)00763000284237(17)280415(10)D033035
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
P100018
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/28/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPED2-375-12
Device Catalogue NumberPED2-375-12
Device Lot NumberD033035
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/25/2025
Supplement Date Received by Manufacturer07/09/2025
Initial Report FDA Received Date06/30/2025
Supplement Report FDA Received Date07/28/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/16/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-