• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SPECTRUM VASCULAR BIOFLO PICC 3F SL-55CM MST-70 KIT VALVED PG; PERIPHERALLY INSERTED CENTRAL CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SPECTRUM VASCULAR BIOFLO PICC 3F SL-55CM MST-70 KIT VALVED PG; PERIPHERALLY INSERTED CENTRAL CATHETER Back to Search Results
Model Number H965458020
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/03/2025
Type of Reportable Event Malfunction
Event or Problem Description
An end user experienced an issue when using a bioflo picc 3f sl-55cm mst-70 kit valved pg.Post placement, it was noted the hub was cracked.The picc required removal.The patient did not experience any adverse effects or harm because of this incident.
 
Additional Manufacturer Narrative
The device has been returned and is currently under evaluation.This report will be updated with the results of the investigation in a follow-up report.Reference (b)(4).
 
Additional Manufacturer Narrative
This report is an update to report 3030714660-2025-00006.Returned for evaluation was one (1) bioflo picc.As received, the female luer lock component of the purple valve was confirmed to be cracked.There were no other visual defects noted to the valve.The customer's complaint description is confirmed for hub cracked based on evaluation of the returned sample complaint.No component molding non-conformances or other damage was observed during visual inspection of the returned sample.A dhr review of the packaging/assembly/valve lots were reviewed for any deviations related to the reported hub crack failure mode.The review confirms that the lots met all material, assembly and performance specifications with no internal non- conformance reports (ncr) written.The most likely root cause for the cracked female valve housing is due to an over tightened connection with a mating male luer (likely a needleless connector) and/or inadequate flushing.A contributing factor to the over-tightened connection may be the mechanics/hand placement during infusion access.It is preferable to grasp the needleless connector (nc) when connecting a syringe/tubing to it rather than grasping the luer hub while making a connection with the nc.Grasping the luer hub while connecting a syringe/tubing set to the nc can transmit additional torque to the female luer hub.Inadequate flushing of the device lumen may also be a contributing factor.An instructional video regarding the care and maintenance of the bioflo picc has been created; reference youtube video, "angiodynamics bioflo picc care and maintenance instructional video" uploaded on november 18, 2020.This video is intended as a visual representation of information provided in the dfu for clinicians who are responsible for the care and maintenance of piccs but are not responsible for picc placement.Labeling review: the dfu that is supplied with the device (ifu: 16600224-01) contains the following precaution: it is recommended that only luer lock accessories be used with the bioflo¿ picc with endexo¿ and pasv¿ valve technology.Repeated over-tightening may reduce hub connector life.Do not use hemostats to secure or remove devices with luer lock hub connections.Flushing - recommended procedure: 1.Flush the catheter after every use, or at least every seven days when not in use, to maintain patency.Use a 10 ml syringe or larger.2.Flush the catheter with a minimum of 10 ml of sterile normal saline, using a "pulse" or "stop/start" technique.Warning: if using bacteriostatic saline, do not exceed 30ml in a 24-hour period.3.Disconnect the syringe and attach a sterile end cap to each luer lock hub.Note: this is the recommended flush procedure for this catheter.If using a different procedure than listed above, the use of heparin may be necessary.Follow institutional protocol for catheter flushing.Precaution: incompatible drug delivery within the same lumen may cause precipitation.Ensure that the catheter lumen is flushed following each infusion.Precaution: if resistance is met when flushing, it is recommended that no further attempts be made.Further flushing may result in catheter rupture.Refer to institutional protocol for clearing occluded catheters.Precaution: place a cap on the hub after use to reduce the risk of contamination.A review of similar complaints noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOFLO PICC 3F SL-55CM MST-70 KIT VALVED PG
Common Device Name
PERIPHERALLY INSERTED CENTRAL CATHETER
Manufacturer (Section D)
SPECTRUM VASCULAR
50 main st
ste. 1000
white plains NY 10606
Manufacturer (Section G)
ANGIODYNAMICS
603 queensbury ave
queensbury NY 12804
Manufacturer Contact
james solorzano
50 main st
ste. 1000
white plains, NY 10606
MDR Report Key22366445
Report Number3030714660-2025-00006
Device Sequence Number18512279
Product Code LJS
UDI-Device IdentifierH965458020
UDI-PublicH965458020
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K150448
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 09/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberH965458020
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/03/2025
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date06/30/2025
Supplement Report FDA Received Date09/10/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-