It was reported that the patient was admitted to the hospital on (b)(6) 2025 due to scrotal swelling and fever for 3 days.After admission, they developed urinary retention and had a urinary foley catheter was placed on the same day.During the morning rounds on (b)(6), it was found that the drainage of the urinary catheter was not smooth, and the physical examination found that the urinary catheter had fallen off.The urinary catheter was replaced, and no harm was caused to the patient.
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
It was reported that the patient was admitted to the hospital on (b)(6) 2025 due to scrotal swelling and fever for 3 days.After admission, they developed urinary retention and had a urinary foley catheter was placed on the same day.During the morning rounds on june19, it was found that the drainage of the urinary catheter was not smooth, and the physical examination found that the urinary catheter had fallen off.The urinary catheter was replaced, and no harm was caused to the patient.
|
|
The reported event was inconclusive because no sample was returned.It was unknown whether the device had met relevant specifications.A potential root cause for this failure mode could be user related (example: contact with sharp object)/mechanical failure/operator error/thin rubberize layer).A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.The instructions for use were found adequate and states the following: structure & composition: the product include three series: bardex, bardic and bardia.The composi-tion of each series is as follows: bardex, bardic and bardia bi-lumen series are composed of catheter body, inflation cone interface, balloon and valve.Bardia tri-lumen is composed to tube body, infla-tion cone interface, deflation cone interface, flush cone interface, balloon and valve.¿the product is with silicone coating.The material of catheter body, inflation conical interface, deflation conical inter-face, flushing conical interface, balloon is natural latex.The material of valve is polypropylene.Steri-lized by gamma radiation.Disposable.Scope of application: foley catheter is intended for use in the drainage of urine from the bladder of children and adults.Precaution: this product contains natural rubber latex which may cause allergic reaction.Sterile un-less package has been opened or damage.Warning: do not use petroleum substrate lubricants with the latex- based urethral catheters such as petroleum jelly and label liquid paraffin, which will damage latex and may burst balloon.Water-soluble lubricant can be used.Valve type: use luer slip syringe.Do not use needle.To deflate catheter balloon: gentle insert a syringe in the catheter valve.Never use excessive force to make the syringe ¿stick¿ in the valve.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen.If permitted by hospital practical, the valve may be cut off.If this fails, please contact an adequately trained professional for assistance, as directed by hospital practical.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.Storage: catheters need to be stored at room temperature and keep away from the direct light and preferably stored in the original packing box.Do not use the product of have latex allergy.Correction- d.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|