| Catalog Number |
ILPC343U |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The doctor reports that item ilpc343u fractured at the screw area and was removed procedure was concluded.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimvie complaint number (b)(4).A1: patient identifier: unknown / not provided.A2: age at time of the event: unknown / not provided.A3: gender: unknown / not provided.A4: patient weight: unknown / not provided.B3: date of event: unknown / not provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimvie complaint number cmp-(b)(4).The following fields have been updated: b5: describe event or problem d2: type of the device: corrected g6: type of report h2: follow up type h11: corrected data.
|
| |
|
Event or Problem Description
|
|
No further event information is available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimvie complaint number (b)(4).The following fields have been updated: b4: date of this report b5: describe event or problem g3: date received by manufacturer g6: type of report h1: type of reportable event h2: follow up type h3: device evaluated by manufacturer h6: adverse event problem h10: additional narrative zimvie received one (1) ilpc343u, (certain low profile one-piece abutment 3.4mm(d) x 3mm(h)) for evaluation.Visual evaluation was performed, abutment's screw was observed fractured.Dhr review was completed for the subject lot number 2120030223.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 2120030223 for similar events and no other complaint was identified.Review completed utilizing keywords: dental : functional : fracture : screw.Based on the investigation and risk management file review, the most likely root cause determined from the investigation was torque applied during placement/seating exceeds recommended value.Therefore, based on the available information, a device malfunction did occur.The abutment's screw was observed fractured.The reported event was confirmed.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.
|
| |
|
Event or Problem Description
|
|
No further event information is available at the time of this report.
|
| |
|