Model Number SK12
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Nonunion/Delayed-union Bone Fracture (2369)
Date of Event 06/10/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was reported that after an initial bunion surgery, all hardware was removed and replaced during revision surgery on (b)(6) 2025 due to non-union.During the hardware removal, a broken screw was discovered.No devices were returned for evaluation.Device specific information was not available; therefore, a review of device history records was not able to be performed.However, all non-conformances for possible kits utilized in surgery were reviewed and no non-conformances or issues during the manufacture or release of the products were identified to date that could have contributed to what was reported.A number of factors could have contributed to what was reported and although the most likely cause cannot be determined based on the available information and without the return of the device, the broken screw was likely subjected to excessive bending stresses which resulted in the breakage and contributed to the non-union.The company will supplement the mdr as necessary and appropriate.
 
Event or Problem Description
It was reported that after an initial bunion surgery, all hardware was removed and replaced during revision surgery on (b)(6) 2025 due to non-union.During the hardware removal, a broken screw was discovered.No other patient outcomes were reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAPIPLASTY SYSTEM 1
Common Device Name
PLATE
Manufacturer (Section D)
TREACE MEDICAL CONCEPTS, INC.
100 palmetto park place
ponte vedra FL 32081
Manufacturer (Section G)
TREACE MEDICAL CONCEPTS, INC.
100 palmetto park place
ponte vedra FL 32081
Manufacturer Contact
renae ginter
100 palmetto park place
ponte vedra, FL 
9043735940
MDR Report Key22371011
Report Number3011623994-2025-00069
Device Sequence Number18130812
Product Code HRS
UDI-Device Identifier00810111221337
UDI-Public00810111221337
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K240173
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 07/01/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSK12
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/10/2025
Initial Report FDA Received Date07/01/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-