SMITH & NEPHEW ORTHOPAEDICS LTD HEMI HEAD 46MM; PROSTHESIS, HIP, HEMI-, FEMORAL, METAL Back to Search Results
Catalog Number 74122546
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Foreign Body Reaction (1868); Osteolysis (2377)
Date of Event 02/12/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal reference number: (b)(4).H3,h6 this complaint was opened by smith+nephew to document a patient complication reported that includes reference to the use of a smith+nephew product without evidence about a specific product problem.The reported complication relates to known inherent procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to confirm a relationship between the reported event and the device or identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
 
Event or Problem Description
It was reported that after undergoing a bhrmh in his left hip joint on (b)(6) 2010, patient experienced periprosthetic osteolysis of internal prosthetic.This adverse event was addressed by conducting a revision surgery on (b)(6) 2024, during which, surgeon converted to total hip arthroplasty.They encounter small fluid collection and some adverse soft tissue reaction.This was debrided and irrigated.Abductor tendons were intact without significant involvement.The femoral and acetabular components were solid.Sterile dressing was applied.Postoperative surgical checklist was also performed.Finally, patient was transferred to pacu in good condition.
 
Additional Manufacturer Narrative
H3, h6: as of today, the implanted devices, all of which were used in treatment have not been returned for evaluation.A review of the historical complaints data for the sleeve was performed using batch numbers, part numbers and the reported failure modes to evaluate patterns of repeated failures or defects.No other similar complaints have been identified for the sleeve.Without a definitive batch number, a complete review of the historical complaints data cannot be performed for the hemi head.A review of historical complaints data was performed using the part numbers and the reported failure modes to evaluate patterns of repeated failures or defects.Similar complaints have been identified for the hemi head.This failure will continue to be monitored via routine trending.In the absence of the actual device, the production records were reviewed for the sleeve reportedly involved in this incident.The released sleeve involved met manufacturing specifications upon release for distribution.As no device batch numbers were provided for the hemi head, manufacturing record review could not be performed.The review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.The alleged failure modes and associated risks have been anticipated within the risk file and the anticipated risk level is still adequate.No further actions are required at this time.A review of historic escalation actions related to the products and similar complaint events was performed.Following the review, prior applicable escalation actions were identified and confirmed to reduce associated risks as far as possible.No further escalation actions are required.The available medical documents were reviewed.A clinical root cause could not be determined.The left hip effusions noted on xray and mri are consistent with the intraoperative findings of adverse soft tissue reaction.However, the clinical root cause of the adverse soft tissue reaction cannot be definitively confirmed.A definitive root cause cannot be determined.Specific factors known to contribute to the alleged fault are excessive physical activity levels, unreasonable stress on replacement system, excessive patient weight, trauma to the joint replacement, loosening of components may increase production of wear particles and accelerate damage to the bone.Should the devices or additional information be received, the complaint will be reopened.Based on this investigation, the need for corrective and preventative actions is not indicated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMI HEAD 46MM
Common Device Name
PROSTHESIS, HIP, HEMI-, FEMORAL, METAL
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK  CV31 3HL
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK   CV31 3HL
Manufacturer Contact
holly topping
1450 brooks rd.
memphis, TN 38116
5123913905
MDR Report Key22371728
Report Number3005975929-2025-00066
Device Sequence Number18146796
Product Code KWL
UDI-Device Identifier00885556071588
UDI-Public00885556071588
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/28/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number74122546
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/04/2025
Supplement Date Received by Manufacturer07/24/2025
Initial Report FDA Received Date07/01/2025
Supplement Report FDA Received Date07/28/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
71356109/LOT NUMBER: 09CM02269; 74120154/LOT NUMBER: 088600; 74222200/LOT NUMBER: 10AW25489
Outcome Attributed to Adverse Event Required Intervention;
Patient Age66 YR
Patient SexMale
Patient Weight94 KG
-
-