Model Number TFL-FBX150S
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/06/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the laser fiber of the powered laser surgical instrument broke during a kidney stone laser treatment and appeared burned at both ends.The procedure was completed using a similar device.There were no reports of patient harm.
 
Additional Manufacturer Narrative
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
 
Additional Manufacturer Narrative
The mdr supplemental report is being submitted to provide updated fields and final investigation results.Updated fields: d9, h2, h3, h6, h11.The device was returned for evaluation and the customer's allegation was confirmed.A definitive root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: cause traced to component failure.Olympus will continue to monitor the performance of this device.
 
Event or Problem Description
No additional information received from the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
150 MICRON TFL SINGLE USE FIBER
Common Device Name
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
GYRUS ACMI, INC.
9600 louisiana avenue north
brooklyn park MN 55445
Manufacturer (Section G)
GYRUS ACMI, INC.
9600 louisiana avenue north
brooklyn park MN 55445
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key22372798
Report Number3011050570-2025-00667
Device Sequence Number8226443
Product Code GEX
UDI-Device Identifier00821925043886
UDI-Public00821925043886
Combination Product (Y/N)N
Initial Reporter CountrySW
PMA/510(K) Number
K183647
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 11/20/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTFL-FBX150S
Device Lot NumberKR450008
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 06/11/2025
Supplement Date Received by Manufacturer11/04/2025
Initial Report FDA Received Date07/01/2025
Supplement Report FDA Received Date11/20/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-