• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: REMOTE DIAGNOSTIC TECHNOLOGIES LTD. TEMPUS PRO; MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

REMOTE DIAGNOSTIC TECHNOLOGIES LTD. TEMPUS PRO; MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS) Back to Search Results
Model Number 00-1026-R
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/18/2025
Type of Reportable Event Malfunction
Event or Problem Description
This report is based on information provided by service engineer and has been investigated by the philips complaint handling team.Philips received a complaint on the tempus pro indicating the dc (direct current) connector is extremely loose.There was no patient involvement and no impact alleged as this was a bench finding.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TEMPUS PRO
Common Device Name
MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)
Manufacturer (Section D)
REMOTE DIAGNOSTIC TECHNOLOGIES LTD.
ascent 1, aerospace centre
aerospace boulevard
farnborough GU14 6XW
UK  GU14 6XW
Manufacturer (Section G)
REMOTE DIAGNOSTIC TECHNOLOGIES LTD.
ascent 1, aerospace centre
aerospace boulevard
farnborough GU14 6XW
UK   GU14 6XW
Manufacturer Contact
sharunyan thavarajan
ascent 1, aerospace centre
aerospace boulevard
farnborough GU14 -6XW
UK   GU14 6XW
MDR Report Key22372903
Report Number3003832357-2025-000502
Device Sequence Number8016956
Product Code MHX
UDI-Device Identifier05060472441058
UDI-Public(01)05060472441058
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K201746
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/01/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number00-1026-R
Device Catalogue Number989706000101
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/28/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/18/2025
Initial Report FDA Received Date07/01/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-