ETHICON INC. SFX SPI PDS+ BI VIO 18IN2 USP1 D/A CTX; SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE
|
Back to Search Results |
|
| Catalog Number |
SXPP2B436 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/04/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a patient underwent a c-section procedure on (b)(6) 2025 and barbed suture was used.During the procedure, according to the gynecologic surgeon, she performed a cesarean section with a resident and they used the suture (b)(6) for uterus closure.During the suturing process, while the resident sutured the first layer of the uterus and made the reverse suturing (backstitch), she pulled the suture, the needle detached from the suture and her hand swung towards her face (the needle did not touch her face).In order to complete the uterus closure ,she used a vicryl suture and completed the uterus closure successfully.No adverse patient consequences were reported.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # = >(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested, and the following was obtained via: ¿ please clarify, did this issue contribute to any patient adverse event? no.- please clarify, did this issue contribute to any patient adverse event? no - could you confirm if this complaint has been reported twice: (b)(4) and (b)(4)? please note that (b)(4) , (b)(4) and (b)(4) were reported by the same surgeon on 3 different surgeries.- please provide the source or name and title of external person providing answers to follow-up (external person submitting answers to sales rep or loc/bq) hagit lahat additional information: was surgery delayed due to the reported event? --> unknown, was procedure successfully completed? --> unknown, were fragments generated? --> unknown, if yes, were they removed easily without additional intervention? --> unknown, patient status/ outcome / consequences --> , was other medical intervention (e.G.X-rays, additional procedures, prescriptions, otc, revision) required: --> unknown, is the patient part of a clinical study --> unknown, (b)(6).Device property of -->none, device in possession of -->none.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).H6 component code: g07002 - no device problem found.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information: h6, d9, h3.H3 investigation summary: the product was returned to ethicon for evaluation.Two representative unopened samples were received for analysis.Product code sxpp2b436.The complaint sample was not received for evaluation.Upon initial inspection of the samples, no external damages were observed on the external packets.In order to evaluate the conditions of the returned samples, the packets were opened, and no defects were detected.The swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along the strand to detect any issue related and no defects were observed during evaluation.A functional test was performed using instron equipment, and the pull force results were above the minimum requirements.The event described could not be confirmed as the device performed without any defect noted.It should be noted that as part of our quality process, each batch is randomly inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.The event described could not be confirmed as no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Corrected data: d1 product code.
|
| |
|
Search Alerts/Recalls
|
|
|