| Model Number |
PS-001402 |
| Medical Device Problem Code |
False Negative Result (1225)
|
| Health Effect - Clinical Code |
Sexually Transmitted Infection (4555)
|
| Date of Event |
06/13/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was requested back for further evaluation however the customer advised that the device was not available.Since the device was not returned, the root cause of the reported failure could not be determined.Refer to page 13 and 14 of the instructions for use (ifu) for the list of limitations which contains information on erroneous results (limitations 4-8), and interpreting the test results (limitation 13).There will be no corrective actions taken at this time as the root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Visby will continue to monitor for trends.Submission of this report does not constitute an admission that medical personnel or a health product caused or contributed to an adverse event.C4: treatment/ therapy start and stop dates (use block c4 to record the date that a diagnostic test was performed for reports that involve an in vitro diagnostic product.) start date: (b)(6) 2025, stop date: (b)(6) 2025.
|
| |
|
Event or Problem Description
|
|
Customer reported a discrepant result for a patient's chlamydia status.The visby medical sexual health test was interpreted as negative for chlamydia.The results from an additional test were positive for chlamydia.The vaginal samples for the two tests were collected by the patient on the same day.A third test was not performed.
|
| |
|