• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: US ENDOSCOPY PADLOCK CLIP PRO-SELECT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

US ENDOSCOPY PADLOCK CLIP PRO-SELECT Back to Search Results
Model Number C913131
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/19/2024
Type of Reportable Event Malfunction
Event or Problem Description
The user facility reported that during a patient procedure the padlock clip pro-select defect closure system did not deploy the clip.No report of injury or procedure delay.
 
Additional Manufacturer Narrative
The padlock clip is used for clip placement within the gastrointestinal tract and consists of a single-use clip within a delivery system.The clip component of the device is within a delivery housing which is mounted and secured at the distal tip on the outside surface of a flexible endoscope.A linking cable rests along the outside of the endoscope's insertion tube and connects the delivery housing/clip to a control handle and thumb actuator outside of the patient.The user depresses the thumb actuator to deploy the clip.The device subject of the reported event was not returned to steris for evaluation as the user facility discarded it.An exact cause could not be determined as the device was not returned to steris for evaluation.The dhr for the subject lot was reviewed and confirmed the lot was manufactured according to specifications; no abnormalities were noted.User facility personnel stated that there were no issues noted during their inspection of the device prior to use.Based on our investigation and similar events where the device was evaluated, improper handling by user facility personnel of the device and linking cable which is part of the mechanism to deploy the clip was likely the cause of the reported event.The instructions for use (ifu) include the following statement about proper handling of the linking cable, "avoid bending the linking cable too aggressively or using a grasping/clamping device on the linking cable as it can create a "kink" and/or disable device function." additionally, the ifu includes instructions to inspect the linking cable for kinks prior to use, specifically, "inspect the entire padlock clip defect closure system for kinks in the linking cable or other damage to the delivery housing, control handle body, and the thumb actuator (cracks, breaks, protruding or missing parts).If damage is evident do not use this product and contact your local product specialist or customer service representative." no additional issues have been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PADLOCK CLIP PRO-SELECT
Common Device Name
CLIP
Manufacturer (Section D)
US ENDOSCOPY
5976 heisley road
mentor 44060
Manufacturer (Section G)
US ENDOSCOPY
5976 heisley road
mentor OH 44060
Manufacturer Contact
coletta cohara
5976 heisley road
mentor 44060
4404130648
MDR Report Key22375122
Report Number1528319-2025-00060
Device Sequence Number8025697
Product Code PKL
UDI-Device Identifier00724995184254
UDI-Public(01)00724995184254(11)240603(17)270603(10)10441090
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/01/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberC913131
Device Catalogue NumberC913131
Device Lot Number(10)10441090
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/11/2025
Initial Report FDA Received Date07/01/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/03/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-