MAUDE Adverse Event Report: ZIMMER BIOMET, INC. TPRLC 133 T1 PPS SO 17X154MM; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. TPRLC 133 T1 PPS SO 17X154MM; PROSTHESIS, HIP Back to Search Results
Catalog Number 51-103170
Medical Device Problem Code Tear, Rip or Hole in Device Packaging (2385)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/13/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).G3: japan.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a supplemental mdr will be submitted.
 
Event or Problem Description
It was reported that damaged sterile packaging was identified in the warehouse during the inspection of circulated stock items.No additional information available for the reported event.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: d4; d5; d9; g3; h2; h3; h6.Visual evaluation of the returned product/provided photos identified damage to the sterile packaging (blister).Sterility has not been compromised.The root cause of the reported event can be attributed to transit damage and a packaging design issue.Actions have been previously raised to implement improved packaging changes.These packaging changes will reduce the number of transit related packaging damage events.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor trends.
 
Event or Problem Description
No additional information available for the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TPRLC 133 T1 PPS SO 17X154MM
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key22375615
Report Number0001825034-2025-01936
Device Sequence Number4617544
Product Code LZO
UDI-Device Identifier00880304498341
UDI-Public(01)00880304498341(17)311220(10)7162018
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K200196
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/14/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number51-103170
Device Lot Number7162018
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/13/2025
Supplement Date Received by Manufacturer10/10/2025
Initial Report FDA Received Date07/01/2025
Supplement Report FDA Received Date10/15/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/20/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexUnknown
-
-