LIVANOVA USA, INC. PULSE GEN MODEL 1000-D; GENERATOR Back to Search Results
Model Number 1000-D
Medical Device Problem Codes Fracture (1260); High impedance (1291); Incomplete or Inadequate Connection (4037)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/06/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the patient's physician suspects a possible lead fracture and wants to send in x-ray images to be reviewed.X-ray images have not been reviewed to date by livanova.No known surgical intervention has occurred to date.No other relevant information has been received to date.
 
Event or Problem Description
Additional information received noting that the patient has not experienced any recent trauma or manipulation.
 
Event or Problem Description
Additional information received noting that the patient underwent an exploratory surgery, and the lead was found to be loose.The surgeon tightened the lead and no devices were replaced.
 
Event or Problem Description
Per the physician, the bottom electrode lead was not screwed into place, onto the ipg site.The physician noting "the bottom electrode lead was not screwed into place onto the ipg site" this will be assessed as incomplete pin insertion being the cause of the high impedance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PULSE GEN MODEL 1000-D
Common Device Name
GENERATOR
Manufacturer (Section D)
LIVANOVA USA, INC.
100 cyberonics boulevard
houston TX 77058
Manufacturer (Section G)
HOUSTON
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
cindy scott
100 cyberonics blvd
suite 600
houston, TX 77058
2816672681
MDR Report Key22379500
Report Number1644487-2025-10317
Device Sequence Number8430038
Product Code LYJ
UDI-Device Identifier05425025750528
UDI-Public05425025750528
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
P970003
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 01/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/26/2026
Device Model Number1000-D
Device Lot Number206384
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/06/2025
Supplement Date Received by Manufacturer07/03/2025
08/29/2025
12/26/2025
Initial Report FDA Received Date07/01/2025
Supplement Report FDA Received Date07/29/2025
09/23/2025
01/20/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/20/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Patient Age44 YR
Patient SexFemale
-
-