| Model Number |
8229706 |
| Medical Device Problem Codes |
Image Display Error/Artifact (1304); Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Codes |
Unintended Extubation (4564); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/24/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted as part of remediation activities associated with capa#: 643143, which is addressing mdr reporting guidance from fdas 1995 preamble, which ensures complaints related to corrections and recalls which were previously reported to fda under 21 cfr 806 are now reported as mdrs; this is not a new malfunction/event.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
A health care provider reported via manufacturer representative that the patient underwent bilateral thyroidectomy.This catheter was used to monitor the recurrent laryngeal nerve during the surgery.Before the surgery, the anesthesiologist used the nerve monitoring endotracheal tube.After intubation, the wire was connected to the nerve monitor.There was no response.During the early intubation test, the debugging test has not passed.There was a prompt that the device has failed to pass the test.After many adjustments, but still no response.Replaced it with another catheter, it could be used normally, and there was no adverse event and no patient injury.
|
| |
|
Search Alerts/Recalls
|
|
|