| Model Number |
453968 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/15/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Combination product: yes.
|
| |
|
Event or Problem Description
|
|
The orsiro mission stent was unable to cross the severely calcified lesion in the lad.When the device was removed, it was discovered that the stent was no longer on the delivery system.The stent was found and left in the left iliac artery.No further treatment was conducted.
|
| |
|
Additional Manufacturer Narrative
|
|
Combination product: yes.Neither the affected device nor the angiographic material was returned.Therefore, no technical investigation on the subject could be performed.The production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.Review of the production documentation confirmed that the device was manufactured according to specifications and passed all in-process and final inspections.During final inspection, every stent system undergoes visual inspection to ensure correct stent embedding and homogeneous crimping of the stent.Further, the stent retention force of a defined amount of samples is tested by means of manufacturing process output monitoring.Based on the conducted investigations, no material or manufacturing related root cause could be determined.
|
| |
|
Event or Problem Description
|
|
The orsiro mission stent was unable to cross the severely calcified lesion in the lad.When the device was removed, it was discovered that the stent was no longer on the delivery system.The stent was found and left in the left iliac artery.No further treatment was conducted.
|
| |
|