RESPIRONICS, INC. LIFEVENT EVO2; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number SP2100X26B
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/20/2025
Type of Reportable Event Malfunction
Event or Problem Description
A trilogy evo ventilator was returned to a third-party service center for service.There was no harm or injury reported.During the evaluation of the device at third-party service center, "service required" codes e-147 and e-149 were found in the ventilator's downloaded error log indicating the o2 flow sensor railed to maximum negative value and o2 flow communication ceased and the sensor is reporting the same value.The device's oxygen blending module board assembly required replacement to address the issue.After repair, the device passed final testing and was confirmed to be operating properly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEVENT EVO2
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key22382397
Report Number2518422-2025-107011
Device Sequence Number9417950
Product Code CBK
UDI-Device Identifier00606959057456
UDI-Public(01)00606959057456
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K181166
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/02/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberSP2100X26B
Device Catalogue NumberSP2100X26B
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/04/2025
Initial Report FDA Received Date07/02/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/14/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-