This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11 additional narrative: the actual device was returned for evaluation.The device was evaluated and it was determined that the reported condition was not confirmed.Therefore, the assignable root cause was not determined.However, during evaluation, it was determined that the device component was damaged, would not run - power supply was damaged and would not run - the electrical control unit was damaged.The device also failed pretests for general condition and power-on test.The assignable root cause was traced to component failure due to wear.
|
|
It was reported that while using the universal battery charger device, when switching on the battery charger on/off switch the electrical facility protection would shut down.It was unknown when the event occurred.The user felt a small ant hill in the fingers (tingling sensation like a shock).There were no reports of injuries, medical intervention or prolonged hospitalization.The date of event was unknown but was known to have occurred in 2025.All available information has been disclosed.
|