| Catalog Number |
68-210 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
11/22/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side "partial deflation".Healthcare professional later reported left side "particles in valve" observed during surgery.Device was explanted and replaced.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side "partial deflation".Healthcare professional later reported left side "particles in valve" observed during surgery.Device was explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: d9, h3, h6.Device evaluation: based on the product analysis performed, the assessments of the complaint codes are: ¿ deflation: not observed.¿ particles in valve: observed.As per the investigation procedure, creases, was observed and none of the observations are found to be potentially related to the manufacturing process, no further actions are required.
|
| |
|