| Catalog Number |
33680004 |
| Medical Device Problem Codes |
Osseointegration Problem (3003); Implant subsidence (4078)
|
| Health Effect - Clinical Codes |
Pain (1994); Inadequate Osseointegration (2646)
|
| Date of Event |
06/04/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Once the investigation is completed, any additional information will be reported in a supplemental report.
|
| |
|
Event or Problem Description
|
|
The digital stryker prophecy team received a ct scan indicating that the patient may require a revision surgery because the tibial component is loose and there is anterior subsidence.The patient is also complaining of pain.The talar implant will be revised to match the tibial inbone component.
|
| |
|
Event or Problem Description
|
|
The digital stryker prophecy team received a ct scan indicating that the patient may require a revision surgery because the tibial component is loose and there is anterior subsidence.The patient is also complaining of pain.The talar implant will be revised to match the tibial inbone component.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could be confirmed since images of ct scans were provided and shows loosening and subsidence of the tibial component.Upon further investigation of the ct scans by healthcare professionals, the following was observed: loosening and anterior subsidence of the tibial tray.Failure of total ankle replacement tar over time is a known complication.In case of a planned revision procedure a medical opinion aiming to determine the root cause of the failure is part of the internal investigation process.Sufficient radiological and clinical information must be provided to enable a meaningful clinical assessment and to identify possible causes of failure.In cases where a ct scan is the only available clinical source, the assessment is limited.No conclusive statement can be provided, because key clinical information e.G.Clinical status, exact symptoms and range of motion of the patient, assessment of the treating physician is missing.The root causes of radiographic findings such as radiolucent areas and bone cysts are often complex and multifactorial and cannot be determined scientifically without this decisive evidence.Due to these limitation, we are unable in such cases to provide statements about the patient, the procedure and/or the device in relation to cause of the failure.Although the issue is most likely patient related, the root cause could not be conclusively determined.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.
|
| |
|