• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GLAUKOS CORPORATION ISTENT INFINITE TRABECULAR MICRO-BYPASS SYSTEM; AQUEOUS SHUNT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GLAUKOS CORPORATION ISTENT INFINITE TRABECULAR MICRO-BYPASS SYSTEM; AQUEOUS SHUNT Back to Search Results
Model Number IS3-US
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/03/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information: the device was not available; therefore, product testing on the actual device could not be performed.The device history record review of the manufacturing lot was performed and there were no non-conformities found to be related to the reported event.A complaint history review was completed.The electronic complaint system was searched for the specified lot#.There were no similar issues reported for this lot #.A review of the device labeling including the risk analysis was completed.The reported issue (trocar fragmentation) is identified to be an anticipated hazard.The residual risk for each potential harm was found to be within the insignificant risk region.The ifu adequately provides instructions for stent implantation, precautions, and warnings for the proper use and handling of the device.Mfr# reference: (b)(4).
 
Event or Problem Description
It was reported that the trocar broke during attempt to implant the stents from a trabecular microbypass stent system.Per report, the trocar fragments were removed intraoperatively from the patient's operative eye using forceps, and a back-up device was used to complete the procedure with no report of patient injury.Additional information has been requested.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ISTENT INFINITE TRABECULAR MICRO-BYPASS SYSTEM
Common Device Name
AQUEOUS SHUNT
Manufacturer (Section D)
GLAUKOS CORPORATION
229 avenida fabricante
san clemente CA 92672
Manufacturer (Section G)
GLAUKOS CORPORATION
229 avenida fabricante
san clemente CA 92672
Manufacturer Contact
brittany massey
229 avenida fabricante
san clemente, CA 92672
949367960
MDR Report Key22384435
Report Number2032546-2025-00071
Device Sequence Number8818231
Product Code KYF
UDI-Device Identifier00853704002449
UDI-Public01008537040024491032043417260430
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K220032
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/02/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2026
Device Model NumberIS3-US
Device Lot Number320434
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/17/2025
Initial Report FDA Received Date07/02/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/13/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-