• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL FAST-CATH¿ TRANSSEPTAL GUIDING INTRODUCER SWARTZ¿ SL TRANSSEPTAL SERIES, SL1¿ 8.; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL FAST-CATH¿ TRANSSEPTAL GUIDING INTRODUCER SWARTZ¿ SL TRANSSEPTAL SERIES, SL1¿ 8.; INTRODUCER, CATHETER Back to Search Results
Model Number 406849
Medical Device Problem Codes Material Puncture/Hole (1504); Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/14/2025
Type of Reportable Event Malfunction
Event or Problem Description
During a supraventricular tachycardia procedure, the sheath was punctured when attempting to perform transseptal puncture.Puncture was attempted using a sheath, interatrial septum heartbeat, and needle.Repeated attempts did not break through the interatrial septum.It was found that the heart rate was a distorted and the sheath had a puncture.The needle was exchanged, and the interatrial septum was then crossed.The procedure was successfully completed with no adverse patient consequences.
 
Additional Manufacturer Narrative
One 8.5f fast-cath introducer sheath and dilator were received for evaluation.The dilator had been perforated proximal to the distal tip.A needle/stylet assembly from current inventory was inserted and advanced through the dilator/sheath assembly; however, the needle could not advance past the location of the perforation.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The swartz/fast-cath transseptal guiding introducer instructions for use (ifu) states, ¿during insertion, always use the stylet to facilitate needle passage through the dilator/sheath assembly.Failure to use the stylet may inhibit advancement of the needle and may result in inadvertent needle puncture of the dilator/sheath assembly or skiving of material from the inner surface of the dilator.¿ the cause of the perforation is consistent with not following the instructions for use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FAST-CATH¿ TRANSSEPTAL GUIDING INTRODUCER SWARTZ¿ SL TRANSSEPTAL SERIES, SL1¿ 8.
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
ABBOTT MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
wendy mccarty
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key22384436
Report Number3005334138-2025-00413
Device Sequence Number8818232
Product Code DYB
UDI-Device Identifier05414734203388
UDI-Public(01)05414734203388(10)10106137(17)261031
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K061015
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/28/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number406849
Device Catalogue Number406849
Device Lot Number10106137
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/14/2025
Supplement Date Received by Manufacturer07/21/2025
Initial Report FDA Received Date07/02/2025
Supplement Report FDA Received Date07/28/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/01/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-