(b)(4).Date sent: 07/02/2025.D4: batch #213d27.Additional information was requested, and the following was obtained: 1.What part of the device was broken? branches or cartridge.It wasn¿t able to fire it.2.Did any piece's fall into the patient? no.3.Is the current patient status known? yes, the patient is alright.4.Was there any patient consequence or change in the post-operative care of the patient as a result of the event? no.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the long60a device was received for analysis with no apparent damage, articulated right and with a gst60d reload loaded on the device.The reload was received partially fired 1/5.The device was tested for functionality in the articulated position with the returned reload by resetting and reloading it into the device.The device achieved its complete stroke firing sequence without any difficulties.The staple line and cut line were complete and the remaining staples met the staple form release criteria.The event reported was confirmed and it is related to improper use of the device.The partially fired is consistent with an incomplete or interrupted cycle.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished #long60a, with lot/batch #c9et5w, and no non-conformances were identified.A manufacturing record evaluation was performed for the finished device batch number 213d27, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|